• Treffer 2 von 2
Zurück zur Trefferliste

Experimental and numerical investigations of a small animal coil for ultra-high field magnetic resonance imaging (7T)

  • The purpose of this work was to develop and investigate a radiofrequency (RF) coil to perform image studies on small animals using the 7T magnetic resonance imaging (MRI) system, installed in the imaging platform in the autopsy room (Portuguese acronym PISA), at the University of Sao Paulo, Brazil, which is the unique 7T MRI scanner installed in South America. Due to a high demand to create new specific coils for this 7T system, it is necessary to carefully assess the distribution of electromagnetic (EM) fields generated by the coils and evaluate the patient/object safety during MRI procedures. To achieve this goal 3D numerical methods were used to design and analyse a 8-rungs transmit/receive linearly driven birdcage coil for small animals. Calculated magnetic field (B 1) distributions generated by the coil were crosschecked with measured results, indicating good confidence in the simulated results.

Metadaten exportieren

Weitere Dienste

Teilen auf Twitter Suche bei Google Scholar
Metadaten
Verfasserangaben:Maíra M. Garcia, Tiago R. Oliveira, Daniel Papoti, Khallil Chaim, Maria C. G. Otaduy, Daniel Erni, Waldemar Zylka
DOI:https://doi.org/10.1515/cdbme-2019-0132
ISSN:2364-5504
Titel des übergeordneten Werkes (Englisch):Current Directions in Biomedical Engineering
Dokumentart:Wissenschaftlicher Artikel
Sprache:Englisch
Datum der Veröffentlichung (online):08.01.2020
Jahr der Erstveröffentlichung:2019
Veröffentlichende Institution:Westfälische Hochschule Gelsenkirchen Bocholt Recklinghausen
Datum der Freischaltung:08.01.2020
Jahrgang:5
Ausgabe / Heft:1
Erste Seite:525
Letzte Seite:528
Fachbereiche / Institute:Fachbereiche / Elektrotechnik und angewandte Naturwissenschaften
Lizenz (Deutsch):License LogoEs gilt das Urheberrechtsgesetz