Upgrade of Bioreactor System Providing Physiological Stimuli
to Engineered Musculoskeletal Tissues
(2017)
A novel central control interface (CCI) is developed to improve the modular bioreactor system with regard to extendability and modifiability in Tissue Engineering (TE) applications. This paper presents the results developed in the project with open-source hardware and the graphical programming system LabVIEW. A new platform independent User Interface was further developed to contribute to the new flexibility of the device.
Design and Development of a Bioreactor System for Mechanical Stimulation of Musculoskeletal Tissue
(2023)
We report on the development of a bioreactor system for mechanical stimulation of musculoskeletal tissues. The ultimate object is to improve the quality of medical treatment following injuries of the enthesis tissue. To this end, the tissue formation process through the effect of mechanical stimulation is investigated. A six-well system was designed, 3D printed and tested. An integrated actuator creates strain by applying a force. A contactless position sensor monitors the travels. An electronic circuit controls the bioreactor using a microcontroller. An IoT platform connects the microcontroller to a smartphone, enabling the user to alter variables, trigger actions and monitor the system. The system was stabilised by implementing two PID controllers and safety measures. The results show that the bioreactor design is suited to execute mechanical stimulation and to investigate the tissue formation and regeneration process …