Filtern
Erscheinungsjahr
Dokumenttyp
- Wissenschaftlicher Artikel (33) (entfernen)
Schlagworte
- Implantat (1)
- Kernspintomografie (1)
- Spondylodese (1)
Institut
- Elektrotechnik und angewandte Naturwissenschaften (33) (entfernen)
The aim of this phantom study is to examine radiation doses of dual- and single-energy computed tomography (DECT and SECT) in the chest and upper abdomen for three different multi-slice CT scanners. A total of 34 CT protocols were examined with the phantom N1 LUNGMAN. Four different CT examination types of different anatomic regions were performed both in single- and dual-energy technique: chest, aorta, pulmonary arteries for suspected pulmonary embolism and liver. Radiation doses were examined for the CT dose index CTDIvol and dose-length product (DLP). Radiation doses of DECT were significantly higher than doses for SECT. In terms of CTDIvol, radiation doses were 1.1–3.2 times higher, and in terms of DLP, these were 1.1–3.8 times higher for DECT compared with SECT. The third-generation dual-source CT applied the lowest dose in 7 of 15 different examination types of different anatomic regions.
Cone-Beam computed tomography (CBCT) has become the most important component of modern radiotherapy for positioning tumor patients directly before treatment. In this work we investigate alternations to standard acquisition protocol, called preset, for patients with a tumor in the thoracic region. The effects of the changed acquisition parameters on the image quality are evaluated using the Catphan Phantom and the image analysis software Smári. The weighted CT dose index (CTDIW) is determined in each case and the effects of the different acquisition protocols on the patient dose are classified accordingly. Additionally, the clinical suitability of alternative presets is tested by investigating correctness of image registration using the CIRS thorax phantom. The results show that a significant dose reduction can be achieved. It can be reduced by 51% for a full rotation by adjusting the gantry speed.
Design and Development of a Bioreactor System for Mechanical Stimulation of Musculoskeletal Tissue
(2023)
We report on the development of a bioreactor system for mechanical stimulation of musculoskeletal tissues. The ultimate object is to improve the quality of medical treatment following injuries of the enthesis tissue. To this end, the tissue formation process through the effect of mechanical stimulation is investigated. A six-well system was designed, 3D printed and tested. An integrated actuator creates strain by applying a force. A contactless position sensor monitors the travels. An electronic circuit controls the bioreactor using a microcontroller. An IoT platform connects the microcontroller to a smartphone, enabling the user to alter variables, trigger actions and monitor the system. The system was stabilised by implementing two PID controllers and safety measures. The results show that the bioreactor design is suited to execute mechanical stimulation and to investigate the tissue formation and regeneration process …