We report on the suitability of two different ranges of Hounsfield units (HU) in computed tomography (CT) for the quantification of metallic components of active implantable medical devices (AIMD). The conventional Hounsfield units (CHU) range, which is traditionally used in radiology, is well suited for tissue but suspected inappropriate for metallic materials. Precise HU values are notably beneficial in radiotherapy (RT) for accurate dose calculations, thus for the safety of patient carrying implants. Some of today’s CT machines offers an extended Hounsfield units (EHU) range. This study presents CT acquisitions of a water phantom containing various metallic discs and an implantable-cardioverter defibrillator (IPG). We show that the comparison of HU values at EHU and CHU ranges clearly reveals the superiority and accuracy of EHU. Some geometrical discrepancies perpendicular to slices are observed. At EHU metal artifact reduction algorithms (MAR) underestimates HU values rendering MAR potentially inappropriate for RT.
We report on investigations that illustrate the interaction between the specific immune system and a young avascular tumor growing due to a diffusive nutrient supply. We formulate a hybrid cellular automata-partial differential equation (CA-PDE) model which includes cell cycle dynamics and allows for tracking the spatial and temporal evolution of this elaborate biological system. We present results of two dimensional numerical simulations that, specifically in this work, include special cases of the spherical and papillary tumor growth, the infiltration of immune system cells into the tumor and the escape of tumor cells from the regime of the immune cells.
In this paper, the effect of computed tomography (CT) values of metals in 12-bit and 16-bit extended Hounsfield Unit (EHU) scale on dose calculations in radiotherapy treatment planning systems (TPS) were quantified. Dose simulations for metals in water environment were performed with the software PRIMO in 6MV photon mode. The depth dose profiles were analysed and the relative dose differences between the metals determined with 12-bit and 16-bit CT imaging, respectively, were calculated. Maximum dose differences of ΔAl= 3.0%, ΔTi= 4.5%, ΔCr= 6.2% and ΔCu= 11.6% were measured. In order to increase the accuracy of dose calculation on patients with implants, CT imaging in the EHU scale is recommended.